BIOMEDICAL PHYSICS
11.1. Physics and medicine
- overview, general intro, what goals
(understand, diagnose, treat)
11.2. Scaling
Area
scales
Let us use a quantity L = the "linear
dimension", some measure of how big an object (or an animal, or some
part of it) is. We do not now care about exactly what shape the object has, nor
whether we are measuring the length, width, height, diameter, radius or other
such quantity of it. We will here focus on issues independent of that.
By an area scale A we mean the relation between
some area of two objects of the same shape but different linear dimension L.
For example if the the side of a square is L then its area is A = L2
so if one square has twice the side length of another, it will have four times
its are: A1/A2 = L12/L22.
But the same would be true for a circle A with twice the radius of another; if
one has the radius L1 and the other L2 then
and if their diameters had been L1
and L2 we would also have
Volume scales
In a similar way, the volume of any
three-dimensional geometrical body is proportional to its linear scale cubed,
e.g.
Physical properties which depend on A
This is related to the general formula for heat
transport by conduction through a material, DQ/Dt = -kADT/Dx where DQ = amount of conducted thermal
energy, Dt = time, DT = temperature difference between
the hot and cold end or surface of the material, Dx = the length of the object through
which heat is conducted or the thickness of the surface through which it moves,
k = thermal conductivity (a material constant, low for good thermal insulators)
and A = the area of the surface or a cross section of the object. (This formula
is no longer required in the IB's Thermal physics. The dependency on the area
can also be related to the L = sAT4 formula in
Astrophysics which is generally valid for radiation of heat as well as for
light: L = the power in watts, A = the area of the radiating surface, s = the Stefan-Boltzmann constant. For other
than "black" bodies the formula can be completed by multiplication
with a unitless constant, emissivity, which is 1 for a black body and smaller
for others.
Similar to pressure, but relevant to solid
objects.
[The stress can be tensile stress, if a force is
pulling the object from its ends (as the force of tension in a rope),
compressive stress (the object is being compressed) or shear stress (the force
is acting parallel to the chosen cross section A).]
Physical properties which depend on V
Absolute and relative quantities
An absolute quantity is one which has a
direct dependency on a the linear dimension L (ex. surface area) while a relative
quantity is the ratio or product of two or more absolute quantities (surface
area per unit body mass, heat loss rate per unit mass) or some other quantity
related to one dependent on L (oxygen absorption rate per unit mass).
Consequences for animals
This explains why an elephant does not look
like a scaled-up version of a mosquito. The mass of an animal is, assuming that
animal tissue is mostly made up of water with about the same density,
proportional to volume, and therefore to L3. Bone strength on the
other hand, is proportional to bone cross-section area and therefore to L2.
So the elephants legs must be much thicker compared to the overall size of the
animal than the mosquito's.
Fig: Elephant and mosquito.
Different types of forces may also be important
on different scales. Surface tension (caused by forces between water molecules)
is important for insects (they may walk on water or be trapped inside a drop of
water) while they are of little importance to larger animals.
11.3. Biomedical mechanics
Centre of mass (or gravity)
The center of mass is a point in a body such
that all the forces of gravity acting on the atoms of it can be assumed to act
in that one point for the purposes of translational and rotational mechanical
problems. For homogenous, simple objects the CM- or CG-point is in the geometric
center.
[If the body is not large enough for the
gravity constant g to be different in different parts of it, then CM and CG
would be the same. For living beings this comlplication can be ignored; the
situation is different for astronomical bodies where a tidal effect can be
observed]
Experimentally, the CG of a rigid body can be
found by hanging it in a pivot point around which it can rotate freely and
drawing a vertical line downwards from it, and then repeat this from another
pivot point. Where the lines intersect the CG point is.
[Quantitatively, the x-coordinate of CM and in
practice the CG can be found in any chosen x-dimension from the formula
xCM = (x1 +
x2 + x3 + ....)/(m1 + m2 +m3
+....) = (x1 + x2
+ x3 + ....)/mtot
where xi and mi are the
x-coordinates and masses of the particles of the body. Corresponding formulas
give the y- and z-coordinates]
Forces and torques in the human body
Recall from mechanics the mechanical
equilibrium conditions:
In the elbow joint, the triceps muscle with its
tendons acts with a force F at an attachment point at some distance r from the
joint around which the lower arm can rotate. The resulting torque t = Fr and is therefore greater the bigger r is. For
humans r is rather small, wherefore we are weaker than some monkeys with less
muscle mass but a higher r in the elbow joint.
Fig: Force and torque in an elbow joint
When lifting a heavy object in a
forward-bending position, the upper body rotates around the pelvic joint which
acts as a pivot point. The force of gravity ("load") on the lifted
object acts downwards at the shoulders. The back muscles are attached very near
the pivot point giving them a lower r-value than the load force. To keep the
clockwise and anticlockwise torques constant (which for slow lifting is
approximately the case) there must be a much larger force in the back muscles.
This can lead to injuries, and it is better to lift with the back in a more
vertical position (with bent legs) since that decreases the r for the load and
thereforce the force in the back.
Fig: Lifting objects and force in the muscles
of the back
Lever systems in the human body
Recall from Mechanics that torque is
t = Fr sin q
for the force F acting at the distance r from a
pivot point. If the angle between the force and the line from where it acts and
the pivot is 90o, we can write
t = Fr
The torque supplied by a given force is
therefore larger the longer "arm" it acts on. It is therefore
possible to lift a heavy stone with an iron bar if the pivot is arranged so
that the force of gravity on the stone gets a shorter r than the lifting force.
This method of lifting is called a lever system. There are different
types of levers, where the pivot may be placed in different places relative to
the lifting force (the "effort") and the force of gravity on the object to be
lifted, the "load".
Fig. Heavy stone lifted with bar, pivot near
the heavy stone.
The ratio between the forces called load (Fload)
and effort (Feffort) is defined as the
Mechanical Advantage = load/effort [DB p. 11]
From Mechanics we have that
If the object is lifted at a constant (angular)
velocity, we can approximately write
The distance moved in a circular path by the
points where the forces act is the arc of a circular sector:
Fig: Circular sector, radius r, arc length s,
angle j
The distance s moved is related to the r as s =
2prj/360o or if a is given in radians, s = jr. The angle j must be the same for both load and
effort if the lever is not broken, so the ration between the distances moved
will be:
Dividing this by an arbitrary time t gives a
speed ratio
which here is called the velocity ratio (for a short
t this makes no difference, instantaeneous speed and velocity are the same):
Velocity Ratio = distance moved by
effort/ distance moved by load [DB
p. 11]
From above it is evident that the
Velocity Ratio (VR) = the Mechanical Advantage (MA).
Applications in the human body:
throwing and lifting
The triceps muscle on the back of
the upper arm is attached to a point near the elbow joint, while the force of
gravity on an object in the hand will be further from this pivot. This makes
that lifting things more difficult (MA = reffort/rload
< 1 since rload > reffort), but the gain is that at
the same time VR > 1, and since MA = VR = veffort / vload we also
have vload > veffort. The result is that humans can
throw things (e.g. a stone or a spear)
or hit with an object used as a weapon with a high velocity vload,
even if the speed at which the triceps muscle can contract is limited for
biological reasons.
Fig. Elbow joints and arms on human
and monkey.
On a monkey (chimpanzee?), the
triceps is attached a bit further from the elbow joint, which makes it
"stronger" than a human, but not as good at throwing things. The same
phenomenon can be further developed with tools that increase the rload even
more: clubs for hitting, slings for throwing stones and spear-throwing tools
like an atlatl.
Another application is the known
fact that lifting objects with a bent back puts more stress and a higher risk
for injuries on the muscles in the back than lifting with bent legs and a
straight back:
Fig: Lifting with bent and straight
back
The back muscles are attached close
to the hip joint acting as a pivot, while the arms are attached to the shoulder
much further away. We cannot avoid a situation where rload > reffort, but we
can decrease the needed Feffort by affecting the angle qload :
When lifting with a back bent forward, qload is close to 90o, but when bending
the legs qload is much smaller.
11.4. Biomedical thermal physics
Metabolism
All the time food is being digested in the
stomach and other organs and in addition to various nutrients being utilised,
its chemical energy is turned into thermal energy which varies from the basal
metabolic rate (when sleeping or unconscious) to higher metabolic rates, for
example at physical activity when a lot more thermal power is generated in the
muscles.
Temperature regulation
Humans like most mammals are keeping a rather
constant body temperature, which means that depending on the metabolic rate and
the external circumstances (temperature and others) there may sometimes be an
excess and sometimes a deficiency of thermal energy. Heat may flow into or out
of the body in the same ways as earlier in thermal energy:
Energy and efficiency
We take in energy in the form of food and
expend it to work done (e.g. lifting objects) and waste heat. In this sense the
human body works like an engine, see the Thermal physics topic. E.g. the
efficiency is, as in Mechanics,
e or h = Eout/Ein
or Pout/Pin [not in DB but a similar definition is given
in thermal physics, DB p.6]
11.5. Biomedical waves : Sound and hearing
Intensity
Sound intensity I is defined as
I = P/A [not in DB]
where P = the power transported by a wave and A
the area through which wavefronts (of e.g. sound) progress. (This quantity is
also used in Astrophysics for the light emitted by a star).
The energy of an oscillating particle is
periodically changes from kinetic to elastic potential energy. For an
oscillation of a mass m on a spring with the amplitude A, the energy will be E
= ½kA2, where k = the spring constant. For these oscillations we have
(here given without proof) that
which gives
wherefore the power and also intensity of a
sound wave are proportional to the squares of the frequency and amplitude.
The decibel scale
The ability of the human ear to detect sound
(its loudness) depends on its frequency and the intensity level. The ear is
most sensitive around a frequency of a few thousand Hz, where the lowest
detectable frequency - the "treshold of hearing" is about I0 = 10-12 Wm-2.
A logarithmic scale (similar to the pH-scale in
chemistry and the magnitude scale in Astrophysics) has been constructed, such
that
b = 10 log ( I / I0 ) where
I0 = 10-12 Wm-2 [DB p. 11]
where the sound intensity level in the
dimensionless unit "bel" is log (I/I0)
and b = the intensity level
in decibels, dB.
The ear
This organ consists of and outer, middle and inner
ear. The middle ear transforms sound pressure variations to larger ones in the
fluid in an organ called the cochlea, from which they are converted to nerve
signals sent to the brain.
[More details about the functioning of the ear
are found in many textbooks and here omitted in this version of the compendium]
Audible sounds
The character of the sounds we can hear is
determined by their "pitch" (frequency), their "loudness"
(perceived intensity) and their "timbre" (their characteristic
wave shape, which for most real sounds is not purely sinusoidal but the result
of a superposition of several tones and overtones. This is what makes the same
note played with a flute and a clarinet sound different).
The range of audible frequencies for the human
ear is ca 20 - 20000 Hz, although the upper limit decreases with age. The
threshold of hearing and the perceived loudness depend on the frequency:
Fig. intensity-frequency diagram
Hearing tests and hearing losses
Still to write are these parts of the IB
syllabus:
D.2.10. hearing tests, type and
extent of hearing loss
D.2.11. audiograms with bone and air
conductions curves
D.2.12. conductive and sensory
hearing losses, possible origins
D.2.13. selective freq losses, loss
of speech discrimination
D.2.14. hearing aids
Resistance (impedance) matching in the ear
Let us review some electric circuit theory: Say
that we have a V = 4.5V battery connected to a R1 = 10W resistor. Let this constitute part 1 of the circuit.
If we then connect another resistor R2 in series with R1,
this resistor will be part 2. The first resistor and the voltage we assume to
be constant; the second can be varied. The question is now: what value should
we give R1 so that the power dissipated in it will be maximal?
Try first with R2 = R1 =
10W, which gives Rtot = R1 +
R2 = 20W. So the current (which in a serial
connection is the same in both resistors) is given by Rtot = V/I
=> I = V/Rtot = 4.5 V/ 20W = 0.225A. The power dissipated in R2
is then P2 = R2I2 = 10W*(0.225A)2 = 0.50625 W.
Is this the maximal power?
Try instead with R2 = 5W. Now Rtot = 15W and I = 4.5V/15W = 0.3A. So P2 = (5W*0.3A)2 = 0.45W, which is less than
above.
What about trying with R2 = 15W? Then Rtot = 25W and I = 4.5V/25W = 0.18A. And then P2 =
15W*(0.18A)2 = 0.486W. Also less than
the first attempt.
Trying other values will reveal that R1
= R2 will maximize the power in the second part of the circuit. This
is called resistance matching (or, for AC circuits with capacitors and
solenoids where ordinary resistance is replaced by a similar quantity,
impedance Z = V/I, impedance matching). This can be shown in various
ways; we may combine P2 = R2I2 with I = V/(R1
+R2) to get P2 = V2R2/(R1+R2)2
and make a graph of P2 as a function of R2 for some
constant R1; the graph will have a maximum at R2 = R1.
It can also be shown with calculus as below.
[Calculus-based proof: We have the function y =
ax/(b + x)2 where a = V2 and b = R1 are
constants, and x = R2 the variable. We find the maximum of y(x) by
differentiating it and solving y'(x) = 0. In this we will use the rule that the
derivative of f/g is (f'g-g'f)/g2, here f(x) = ax and g(x) = (b+x)2
= b2 + 2bx + x2 :
Practical applications of this is e.g. building
loudspeaker systems, where maximal power is transmitted to the next part of the
system if its impedance is the same as that of the previous.
But what does all this electric circuit theory
have to do with the functioning of the ear? Well, its all about transferring
waves, that is oscillations or pressure maxima (compressions) in air in the
middle ear to water in the inner ear. This leads to the field of fluid (=
liquid or gas) mechanics. In that field, many theories are very similar to
those in electricity.
The equation R = V/I can be replaced by a
similar one where
It will also be true that the total fluid
resistance for two tubes in series or parallel will follow the same Rtot
= R1 + R2 and 1/Rtot = 1/R1 + 1/R2
formulas as in electricity. And as in electricity, the fluid resistance can be
replaced by a similar quantity, fluid impedance, and to effectively
transmit the power in the sound waves from middle to inner ear we need the same
impedance matching as above. And this will be a big problem, since the
fluid impedance of air and water are very different.
11.6. Ultrasound
Ultrasounds
and infrasounds
By
ultrasounds we mean sounds with a higher frequency than the ca 20000 Hz that a
(young) human can hear. These sounds can be heard or sometimes even produced by
some animals, such as dogs and bats. Sounds with a lower frequency than we can
hear (ca 20 Hz) are called infrasounds and used by elephants to communicate
over long distances.
Ultrasounds
are produced with piezoelectric crystals, which change shape when an electric
potential difference is applied to them, and
therefore
can vibrate and produce sound waves when a high-frequency AC voltage is used.
Ultrasounds in medical use typically have frequencies about 1-10 MHz.
The
SONAR principle (radar with sound)
A
radar (radio detecting and ranging) sends out a pulse of radio waves and
measures the time it takes for it to be reflected back to the radar antenna.
Knowing the speed of the wave, it is then possible to calculate how far away
the target - a ship or an airplane - is. The SONAR ("sound navigation
ranging") or pulse-echo technique is similar for sounds waves, and has
been used since WW2 to find submarines under water. In doing this, the sound pulse
may be reflected or refracted not only by a target submarine or the ocean
floor, but also by layers of water with different temperature and/or salinity.
In
medical use, the ultrasound pulse is (partially) reflected when reaching the
boundary between tissues where sound travels at slightly different speeds,
around the typical value ca 1540 ms-1 for water (which human bodies
mostly consist of). In air this speed is ca 340 ms-1 and in bone ca
4000 ms-1 which gives a very effective reflection, but makes it
difficult to "see" organs behind lungs or bones.
In
order to avoid a strong reflection when the pulse enters from the transducer
into the body (= water), a water-based gel can be used to ensure that the pulse
never travels through air.
A-scans
and B-scans
The
A-scan produces a graph of the received echo intensity as a function of time
since the pulse was sent out. The following pulse is not sent until the first
has gone through the body, and the duration of a pulse is short compared to the
separation in time between them. Pulse echoes (intensity peaks) represent the
passing of the pulse through tissue boundaries
Fig. b06a: Tissues and A-scan graph and B-scan
points (G372)
In a
B-scan, the peaks are represented by points that by a computer are made more
intense, larger, differently coloured or otherwise to represent the height of
the peak in the A-scan. By moving the transducer and/or having several of them
a set of B-scans give a two-dimensional
image of the organs inside the body.
Fig b06b: set of B-scans and image of inner
organ
Factors
affecting the choice of frequency
·
resolution:
the higher the frequency, the shorter the wavelength, and the smaller details
can be observed; e.g. if f = 10 MHz and v = 1500 m/s then limit is given by:
v =
fλ => λ = v / f = 1500 ms-1 / 1000 000 Hz = 0.0015 m = 1.5 mm
·
penetration:
the higher the frequency, the more of the wave is absorbed or scattered by the
water, and this makes it difficult to see organs deeper in the body (compare
this to the scattering of light in the atmosphere: blue light with a high
frequency is scattered more than red with a low frequency, therefore the sky
looks blue in the day, and the sunset red).
11.7. EEG and ECG (not required in the IB)
[ECG
= electrocardiogram: Small electrical voltages are used to detect the action of
muscle cells, especially the heart muscle. Different stages in the operation of
the heart muscle produce different characteristics on a graph of voltage as a
function of time. Malfunctions in the heart are detected as distortions of
these graphs from the normal type
EEG
= electroencefaphalogram: same for brain ]
11.8. Biomedical atomic physics : X-rays
X-ray imaging
Production and properties of X-rays
From Atomic physics, recall that when electrons
in a vacuum tube are accelerated by a p.d or voltage of ca 10000V and then hit
a metal target, where some (ca 1 %) of their kinetic energy is released as X-rays
are produced, and the rest turns into thermal energy. These X-rays are
high-frequency and high-energy (E = hf) photons, with an f higher than UV-rays
but lower than gamma rays. The target material is often tungsten (wolfram) like
the filament in a light bulb, since it has a high melting point. To avoid
overheating of the spot where the stream of accelerated electrons hit, the
target is usually rotated. The X-ray spectrum (graph of intensity as a function
of X-ray frequency or wavelength) has 3 important features:
The area under the graph represents the total
intensity of the X-rays produced.
X-ray quality
The term "X-ray quality" refers to
the intensity of the rays at different wavelengths, and affects the intensity
that actually reaches an organ inside the patient and then the detecting device
(film or electronic detector).
Factors affecting X-ray quality and their
results for features I-III
A. Increasing the potential difference V
I. When more kinetic energy is carried by the electrons, the broad peak of
the continuous curve shifts towards higher f / lower l.
II. The maximum f increases/ minimum l decreases
III. Depending on where the energy levels of the
shells in the target atom are, new characteristic peaks may appear (but all
peaks are at their constant places in the spectrum).
B. Increasing the electron current
More electrons per time are sent through the
X-ray tube (e.g. by increasing the current heating the cathode thus making
thermionic emission of electrons more efficient). The drawback is the heating
of the target material also increases.
I. No change in the shape of the
characteristic curve, but the overall intensity increases
II. No change in the max f / min. l.
III. No change in the position of the
characteristic peaks or in which of them are found, but the general increase in
intensity is noted in them as well.
C. Increasing the atomic number Z of the target
material
For the target atoms, a higher Z makes the
collision between an accelerated electron and target atom more likely to result
in the emission of an X-ray photon than in a heating of the target. The problem
lies in balancing this against the need for target materials with a high
melting point. Common materials are tungsten (wolfram) with Z = 74 and a
melting point = 3370 oC, and platinum with Z = 78 but a melting
point of 1770 oC). More effective cooling (by rotation or otherwise)
of the target opens possibilities for using a higher Z.
I. No change in the shape of the continuous curve, but the overall
intensity values increase.
II. No change in max f. / min l.
III. The characteristic peaks for a different
target atom are different - usually the ones in materials with a higher Z are
found at a higher f/ lower l.
D. Energy selective filters
When EM waves pass through suitable materials,
some wavelengths are absorbed more than others. For visible light, this may
lead to white light changing to coloured light when passing a transparent but
"coloured" material, e.g. coloured glass. There are materials which
have a similar effect on X-rays, absorbing selectively at lower f / higher l and thus improving the resulting X-ray
"quality". For X-rays as for any EM radiation, the term "harder"
refers to a spectrum where higher frequencies are more dominating. Blue light
is "harder" than red or white light.
I. The shape of the continuous curve changes so that its broad peak is
shifted towards higher f / lower l. The overall intensity level and
total area under the curve decreases.
II. No change in max f. / min l.
III. No change in the position of the
characteristic peaks or in which of them are found, but their relative height
may change since they are unevenly affected by the decrease in intensity.
X-ray attenuation
Attenuation means the damping or decrease in
intensity for the X-rays when they pass through the material in the patient's
body. Compare this to light becoming fainter when it passes deeper and deeper
into the ocean.
A. Simple coherent scattering
(and a few words on Compton scattering)
If the X-ray photon an energy E = hf that does
not fit any of the energy level differences in the atom in the patient's
tissue, it may change its direction without losing energy. This will, however,
mean that the intensity of radiation passing through the body in a given
direction (from the X-ray tube to the film or other detecting device)
decreases. This mostly happens in soft tissue but to some extent also in bones.
[Compton scattering occurs when the incoming
photon collides with an electron in such a way that it loses part of its
momentum p = E/c and therefore energy (without losing either speed, which is
always c, or mass, which it does not
have!. This process adds to the decrease in intensity of photons with a given
wavelength since the photon must lose energy in that way]
B. Photoelectric
effect
Recall from Atomic physics that a photon
hitting a target atom may strike out an electron losing all (in proper PE
effect) or part of its energy. The stricken electron may be replaced by another
under emission of a new photon, but mostly with a different, lower, energy. The medical relevance of this is that
since bone tissue contains a different distribution of elements than
surrounding soft tissue (muscles) it is possible to design the X-ray so that
its peak in the intensity curve matches the energy needed for photoelectric
effect in bones. There will then be a relatively greater difference in
attenuation in bones vs. other tissue, and a sharper image of the skeleton can
be produced.
Attenuation
coefficient and intensity
The change (decrease) in intensity ΔI when
radiation passes the distance Δx through materia depends on the intensity
I before passing Δx as:
ΔI = -μIΔx
where μ = a linear attenuation coefficient in the unit m-1. If we turn this
into a differential equation
dI = -μIdx
we will get the solution:
I = I0e-μx DB
p. 11
where I0 = the intensity before
hitting the material and I the intensity at a depth x in it.
[Compare this to radioactive decay in Atomic
physics where the decay probability constant λ in the unit s-1
and the differential equation
dN = - λNdt
gave
N = N0e-lt [DB p.8]
The half-value
thickness
For similar mathematical reasons we will also
get something corresponding to the half-life
T½ = ln2 /
λ [DB p. 8]
namely the half-value
thickness x½ = the depth at which the intensity has been halved:
x½ = ln2 / μ DB
p.11
The attenuation coefficient depends on:
X-ray detection
Basic detection: films and electronics
The X-rays can be detected with certain
photographic films sensitive to them or secondary radiation from intensifying
screens. The developed films are studied agains an illuminated background. The
X-rays can also be detected and recorded by electronic components similar to
those in digital cameras, and displayed on a computer monitor.
X-ray image enhancement
To improve the quality of the images one can:
·
let
the rays pass collimating grid of lead plates, which will
suppress rays not moving in the desired direction
·
move the
collimating grid (or the source and the film) in such a way that the images of
the grid are blurred and those of the patient sharp
·
use
contrast-enhancing substances which make medically interesting
organs more visible when introduced into the patient (barium meal for the
intestines, injected iodine for the blood system)
·
use
intensifying screens, that is materials which when hit by X-rays
release lower-frequency radiation which is more easily registered by the
photographich film/ electronic detector
Computer tomography (CAT scan)
In
an ordinary X-ray image, the rays are sent through a whole part of the
patient's body at one time and the intensity of the rays afterwards redorded
with a film or otherwise. In computerised axial tomography, one could in
principle use just one very narrow ray and one small detector and then move
them in a circle around the patient, although in practice several rays and an
arc-shaped array of detectors are used to produce an image more quickly:
Fig. b08---: Patient (P) with X-ray source and
detector array, ring
The
intensity of one narrow ray - "axis" - that has passed through the patient is only one
number; to produce a two dimensional image of the patient the computer
calculates what the X-ray absorption must be in different parts of the
"slice" of the patient. This problem can be compared to finding what
numbers must be placed in a matrix when the sums of rows, columns and diagonals
are known:
Fig b08----: Box 3 x 3 with numbers 1,1,1;
2,1,7; 1,1,2 and sums including diagonals below 3,4,3,10,4 and on the right
side upwards 4,10,3
To
produce a 3-dimensional image many such thin "slice-images" are made
by moving the source and detector array in a direction perpendicular to the
picture above.
11.9. Other imaging techniques
MRI
or NMR (so-called "magnetic
X-ray")
The
nuclei of atoms are rotating, "spinning", and for those with an odd
mass number (sum of protons and neutrons) such as 1H this makes them act as
small magnets. If the atoms are placed in a strong magnetic field B, then they
can be aligned either in the same or in the opposite direction as the field B.
To
make them switch between these alternatives ("spin up" and "spin
down") requires or releases energy,
which is proportional to the strength of the field B. If the B-field is
directed "up", then energy is released when the nucleus switches from
the opposite direction to one aligned with B:
Fig b09a: B-field , E-levels (G996)
This energy can be absorbed or emitted photons
with the energy E = hf, where the frequency is typically in the radio frequency
(RF) part of the electromagnetic spectrum. The frequency must be such that the
energy of the photon matches the energy absorbed or released in the switching
of the nucleus; this phenomenon is called nuclear magnetic resonance or NMR.
Since
the needed energy and therefore the radio photon frequency depends on the
magnetic field B, one can "excite" (switch to the higher energy
state) nuclei on only one slice of the body by using a magnetic gradient,
that is a B field that grows stronger the further along a chosen direction we
go. By applying another B-field, also with a gradient, perpendicular to that,
we can find out where is this slice an emitted RF photon comes from - that will
be revealed by its frequency. Add to this that the apparatus can be rotated
around the patient and a computer used to find where in the patient we have a
lot of emissions/absorptions, then we can get a very detailed 3-dimensional
image of where in the patient there is more or fewer H-atoms.
This
does, however not give very good medical information about the patient, since
these atoms are present in large numbers everywhere, in water which makes up
most of the body and in all typical organic molecules. What is done then is to
measure how fast the excited H-atoms "decay" back into the lower
energy state; this is to some extent affected by the neighbouring atoms, and therefore
the type of molecule, and consequently the type of tissue we have. This is
called a spin-echo technique.
Of
other atoms present in organic molecules, 12C and 16O are
unsuitable (not having odd nucleon numbers), but 31P has been used
in studies of metabolism (where phosphorus in ADP and ATP molecules are
essential).
Radioactive tracers
Some elements or chemical compounds for
biochemical reasons tend to accumulate in certain organs. For example iodine does
so in the thyroid gland (Sw. sköldkörteln, Fi. kilpirauhanen). If there are
radioactive isotopes of iodine in the environment which may be the case after a
nuclear accident, then one can take an excess of non-radioactive iodine so that
the body cannot take up more of it for some time. Then the accumulation of
radioactive iodine is avoided and the risk of thyroid cancer decreased.
A similar phenomenon can be used for medical
treatment with radioactive tracers.
This means that isotopes of elements which typically accumulate in some organ
but have a short half-life so that they do not contaminate the body for a
longer time are injected, and the uptake in some organ studied. If the uptake
is smaller or bigger than usual it indicates that this organ is not functioning
properly.
The detection is done either by checking blood
or other fluid samples with a Geiger-Muller-detector (or similar) for alpha and
beta emitters which send out radiation that does not easily get out from the
body. For gamma emitters the detection can be done with an external "gamma
camera", a detector for gamma rays. Since gamma rays are electrically
neutral they cannot be found with a GM-tube, but have to be studied indirectly.
A "scintillation detector" uses certain compounds (sodium iodide with
small amounts of tellurium added) where gamma rays can be absorbed and the
subsequently emitted photons of visual light observed.
PET =
positron-emission tomography with coincidence measurements
This is a special case of diagnosis with radioactive
tracers. Recall that there are three types of beta decay: emission of electrons, of their antiparticle positrons, and electron capture. The second of these
types is used, and when then the
positrons are annihilated by
ordinary electrons in the body, two
gamma photons are emitted.
Even if a photon does not have a mass, it does
according to relativity theory have a momentum, and to conserve the momentum in the annihilation event, the photons move
out in opposite directions. (The emitted
positron has a very small momentum and kinetic energy compared to that of the
photons, since all the mass in the positron and electron is converted to energy
as E = mtotc2).
The beta emitter is inserted into the patient
(e.g. as radioactive oxygen 15O inhaled) and a ring of gamma
detectors around e.g. the head used to detect the gamma photons. To distinguish
the relevant photons from others a computer only counts the photon hitting a
detector if another hits the detector exactly opposite to it at the same time (coincidence). In addition the same
imaging techniques as in CAT scans are used to produce a three-dimensional
image of the brain.
11.10. Biomedical nuclear physics
Types of ionising radiation
D.5.1. biol effects of ionizing
radiation
"Dose" units
- mention Bq also
Absorbed dose = Absorbed Energy /
mass [DB
p. 11]
Exposure = total charge / mass [DB p. 11]
Dose equivalent = quality factor x
Absorbed dose [DB p. 11]
D.5.2. exposure, absorbed dose, quality
factor, dose equivalent in dosimetry
Precautions for ionising radiation
There are 3 main ways of reducing the dose of
ionising radiation one gets.
1. Time : the source causes a certain equivalent dose
per time unit, e.g. in microsieverts/hour, and limiting the time of exposure
decreases the total dose received.
2. Shielding : to place material which absorbs a significant
part of the radiation between oneself and the source (Here one must be careful
to avoid secondary radiation emitted by the shielding material when the primary
radiation collides with its atoms, or "Bremsstrahlung", the
electromagnetic radiation emitted whenever a moving charge is accelerated or
decelerated.
3. Distance : for any point source of radiation the
"intensity" of radiation hitting a given area is inversely
proportional to the square of the distance from it
(In Finnish : remember ASE = aika, suoja, etäisyys)
Radiation therapy
D.5.7. radiation therapy for cancer
basics ??
D.5.8. probs: choice of isotope for given
diagnostic or therap.???
Physical, biological and effective half-life
Recall from nuclear physics that the number of
decayed atoms of the N present in a sample in a given time Dt is
DN = - lNDt
where l = the decay constant, related
to the half-life of the nuclide as T½ = ln 2 / l [DB p. 8]. This half-life will
here be called the physical half-life TR (and the decay constant lR) of the radioactive nuclide in
question. But some of the radioactive atoms in the patient's body may also
leave it before they have decayed with exhaled air, urine, feces, vomit, semen
or other ways of losing materia from the body. These processes do not precisely
follow any simple mathematical formula, but a reasonable approximation is that
they are proportional to the number of radioactive nuclei in the body and the
time given, and to some biological decay constant lB. We will then have the total or effective decay constant
1 / TE = 1/TB +
1 / TR [DB
p. 11]